MEMORY HiCORDER MR8880

Number of channels 4 analog channels + 8 logic channels (standard)
Note: Isolated analog channels, isolated input and frame, logic has common GND
Measurement ranges
(10 div full-scale)
4 channels of voltage measurement; mode switchable between instantaneous waveform or RMS value, 10 mV to 100 V/div, 13 ranges, resolution: 1/640 of range
RMS value mode: 30 Hz to 10 kHz, Crest factor: 2
Max. rated voltage Between terminals: 600 V AC/DC, Between terminal to earth: 600 V AC/DC CAT III; 300 V AC/DC CAT IV
Frequency characteristics DC to 100 kHz (±3dB)
Time axis
(High-speed function)
100 µs to 100 ms/div, 10 ranges, Sampling period: 1/100 of range
Recording intervals
(Real-time function)
100 µs to 1 minute, 19 selections (simultaneous sampling in all channels)
Measurement functions High-speed function (high speed recording)
Real-time function (actual time recording)
Memory capacity 14-bits × 1M-words/ch (1 word = 2 bytes)
Removable storage CF card slot ×1 (Up to 2 GB), USB 2.0 memory ×1
Printing [Printer unit is option] 112 mm (4.41 in) × 18 m (59.06 ft), thermal paper roll, Recording speed : 10 mm (0.39 in)/sec
Note: Printing is not supported when using alkaline batteries
Display 5.7-inch VGA-TFT color LCD (640 × 480 dots)
Displayable languages English, Japanese, Chinese
Communication interfaces USB 2.0 mini-B receptacle × 1; Transfers files from the installed CF card or USB memory stick to a PC when connected, and External PC control
Power supply AC adapter Z1002: 100 to 240 V AC (50/60 Hz), 45 VA (include AC adapter, when Real-time recording), 107 VA (include AC adapter, when Real-time recording and printing)
Battery pack Z1000: AC adapter has priority when used in combination with battery pack, recharge with AC adapter 3 hours, Continuous use 3 hours (with back-light ON)
LR6 (AA) alkaline batteries ×8, Continuous use 40 minutes, (with back-light ON, cannot be used with the Printer unit)
DC power supply: 10 to 28 V DC
Dimensions and mass 205 mm (8.07 in)W × 199 mm (7.83 in)H × 67 mm (2.64 in)D, 1.66 kg (58.6 oz) (with the Battery pack installed)
When printer is combined – with main unit: 303 mm (11.93 in)W × 199 mm (7.83 in)H × 67 mm (2.64 in)D, 2.16 kg (76.2 oz) (with the Battery pack installed)
Included accessories Instruction manual ×1, AC adapter Z1002 ×1, Alkaline battery box ×1, Strap ×1, USB cable ×1, Application disk (Wave viewer Wv, Communication commands table) ×1

LabVIEW and MATLAB support

Compatible models LabVIEW MATLAB script
MR8880-20 N/A N/A

Categories , Brand:

Description

Measuring Current Waveforms: How to Use Hioki Memory HiCorders

Learn how to use the Hioki MR8880’s trigger function to definitively record the startup current when a motor is powered on.
Pre-trigger settings let you capture details about the situation before and after the trigger is activated, and numerical data of the peak current waveform can be easily identified using the built-in cursor functions.

Date Sheet

MR8880E20-55M (1)

Reviews

There are no reviews yet.

Be the first to review “MEMORY HiCORDER MR8880”

Your email address will not be published. Required fields are marked *